Minor releases | 810 - xxx

Release 870 | March 2025

Functionalities for Agilent 6495D added

  • Export worklist to MS
  • Intensity threshold in Settings > Quantification > Intensity threshold
  • The "intensity threshold" is used in Quantification > Evaluation > Saturation
    Monitoring, saturation

Release 864 | February 2025

General performance improvements and bug fixes

Release 842 | February 2025

Load results

Fixed issue: Notes linked with a plate run may have caused an error message.

General performance improvements and bug fixes

Release 840 | January 2025

Export worklist to MS | Support of additional autosampler models

Newly supported autosampler models for SCIEX OS.

Autosampler modelRack typePlate type
Agilent Infinity II - 3Multisampler 2H 2H 2H 2H 5Ref96DeepNunc31mm
Agilent Infinity II - 4Multisampler 2H 2B 2B 2B 5Ref96DeepNunc31mm
Agilent Infinity II - 5Multisampler 2B 2B 2B 2B 5Ref96DeepNunc31mm
Shimadzu Nexera 40 DSIL-40 DrawerDeep Well 96

For more information refer to section Export worklist to MS.

Quantification > Evaluation | Issue with "ratio-based" value types

Fixed issues:

  • After validation, values of "ratio-based" value types were displayed as "NA".
  • Error after changing a value type, e.g. from "Concentration" to "Area ratio". Quantification > Evaluation | Values after re-integration
    Fixed issue: After excluding a peak from integration and performing a re-integration, the corresponding values were still displayed.

LIMS > Project | Export worklist to MS

Fixed issue: If a project name contained a comma, an error was shown.

General performance improvements and bug fixes

Release 810 | January 2025

Browser version check

  • A browser version check was implemented.
  • WebIDQ can exclusively been used with supported web browser versions.

Results | Values when a peak was excluded from integration

  • In Quantification > Integration peaks can be excluded from integration, Peak integration options > Remove integration window .
  • If a peak was excluded from integration and for values different from "concentration", see Value type, values may have been displayed. However, values should not have been shown, as no integration was performed.
  • This issue was fixed.

Concentration values were always calculated and shown correctly.

General performance improvements and bug fixes