Skip to main content

Calibration

biocrates kits include an external seven point calibration.

For more information refer to the kit user manual.

Calibration

To calculate concentrations and checking the kit performance, LC data needs to be calibrated. To perform the calibration, upload or load kit data and check the calibration as described below.

Calibration workflow

Recommended workflow | calibration

  1. Check calibration
  2. Calibrate plate run

Check calibration

The table Metabolite shows all metabolites, which are seven point calibrated with a kit. One by one, check the integration parameters for each metabolite.

Example for Ala

  1. In the table Metabolite, select "Ala".
    Select Ala
  2. Check the calibration curve and accuracy of calibration standards in the view Calibration and the tables for calibration standards and QCs on the right.
    Calibration curve and tables

    If the accuracy of a calibration standard is outside the acceptance range, it is highlighted in red.

    Calibration standards can be disabled or included, as described below.

  3. If required, adjust the curve type and weighting type as described in Validation features.

Calibrate plate run

To perform the calibration procedure of the selected plate run, click Calibrate.

Calibrate

The progress is indicated as a blue bar at the upper left edge of the screen.

Progress bar

Next steps, if LC, FIA / FIA XL data was integrated

  • Validation of all plate runs of one kit, click here

Calibration features

The calibration process can be adjusted by one of the following features.

Adjustments can be defined for all metabolites, selected metabolites, or a specific metabolite.

Example "Weighting type"

All metabolitesSelected metabolitesSpecific metabolite
All metabolites selectedSelected metabolites selectedSpecific metabolite selected
All metabolites selectedMultiple metabolites selectedOne metabolite selected
Weighting type - all metabolitesWeighting type - selected metaboliteWeighting type - specific metabolite

Curve type

Define the regression type for all calibration curves of a plate run. These options are available.

  • Linear
  • Linear zero
  • Quadratic

The default and recommended curve type in WebIDQ is Quadratic.

Weighting type

Define the weighting type for all calibration curves of a plate run. These options are available.

  • No weighting
  • 1/X
  • 1/X²

The default and recommended weighting type in WebIDQ is 1/X.

Disable calibration standard level

Calibration standard levels can be disabled from the calibration curve plot or table.

  • Calibration Curve: to disable, click on a calibration standard level point in the plot
  • Table: switch the toggle, active / deactivated (activate / deactivated)

When multiple metabolites are selected, calibration standards of all selected metabolites or a specific metabolites can be activated/deactivated.

Applies to all selected metabolitesApplies to specific metabolite
EnabledActive
Applied to all selected metabolites.
All graphs
Applied to the metabolite, e.g. Ala, shown in the graph.
Active, graph

When multiple metabolites are selected, calibration curves of all selected metabolites are shown.

Calibration curve graphics list

To disable a specific calibration standard level, e.g. level 1 of Arg,

  • Option 1 Click on a calibration level in the calibration curve graphics.
  • Option 2 Disable a calibration level from the list by deactivating the toggle Ala.
    One metabolite inactive

Include a calibration standard level

  • Option 1 Click on a disabled calibration point on the calibration curve plot.
  • Option 2 Enable a disabled calibration level from the list by activating the toggle of one specific metabolite or All metabolites (all selected metabolites).

Well positions and accuracies of calibration standards and QCs are shown in two tables. Accuracy information is updated in real time as calibratoin settings are changed.

Calibration standards table QCs table

Check integration of calibration standards

Click on a calibration standard level, to show the corresponding chromatogram.

Select a calibration standard

Chromatogram

If required, adjust the integration windows and apply the changes by pressing Enter or clicking Reintegrate Reintegrate.