Skip to main content

Integration | features LC

Integrate

Integrate

Integration windows are defined in section Integration workflow LC. The WebIDQ quantification algorithm identifies peaks within defined integration windows. After completing the adjustment of all integration windows of a kit plate, click Integrate.

By defining an integration window, quantification is not performed!

To perform the quantification of a plate run, click Integrate.

The start and end points of peak integrations can be adjusted globally for all integration procedures, available in Settings > Quantification > Integration.

Discard settings

Discard settings

All injection/well specific integration windows (dashed lines) defined by the user within the selected plate run are replaced with the default integration windows.

  1. Select a plate run
    Select a plate run
  2. Click Discard settings

Discard settings replaces all metabolite or sample specific integration windows with their default values (solid lines), defined for injections/wells of a plate run.

Apply as default

Apply as default

Integration windows from a selected injection/well can be stored for further plate run integrations (LC part of a kit).

  1. Select an injection/well from the table Wells, e.g. QC2, well 50.
    Select QC2
  2. Click Apply as default.
  • Integration windows of the selected well, e.g. QC2, are stored for the next integration of a plate run.
  • If no plate run specific integration windows are defined, all default integration windows are replaced with the ones from the selected well.
  • Injection/well specific integration windows are not changed (See discard settings above).

To replace injection/well specific integration windows by windows from one well or a CSV file, see Import integration windows, follow this procedure.

  1. Click Discard settings
  2. Select one well of the kit plate, e.g. QC2, well 50.
  3. Click Import integration windows
  4. While in this example QC2, well 50, is selected, click Apply to plate run

Apply to plate run

Apply to plate run

Integration windows of a selected injection/well can be applied as the integration windows for the whole plate run (LC part of a kit).

  1. Select an injection/well from the table Wells, e.g. QC2, well 50.
    Select QC2
  2. Click Apply to plate run.
  • Integration windows of the selected well, e.g. QC2, are used for all injections/wells of the currently selected plate run.
    Select a plate run
  • Injection/well specific integration windows are not changed (See discard settings above).

Export integration windows

Export integration windows

Integration windows of a selected injection/well can be exported.

  1. Select an injection/well from the table Wells.
  2. Click Export RTs to export the integration windows of a selected sample, e.g. QC2, well 50.
    The export file format is .csv.

Only injection/well specific integration windows can be exported.

Default integration windows cannot be exported.

Import integration windows

Import integration windows

Integration windows can be imported from a WebIDQ compatible .csv file.

Integration windows of a .csv file are applied to a selected injection/well.

  1. Select an injection/well from the table Wells, e.g. QC2, well 50.
    Select QC2
  2. Click Import RTs to apply integration windows to a selected injection/well, e.g. QC2, well 50.

Imported integration windows from a .csv file replace all integration windows of the currently selected injection/well, e.g. QC2, well 50. Integration windows with were defined manually for one sample are not replaced.

Export MS instrument methods - MxP® Quant 1000 and MxQuant kits

Adjusted retention times can be exported directly to instrument methods for MxP® Quant 1000 and MxQuant PITC and NPH methods.

  1. For each method to be exported: In the table “Wells” select a reference sample with updated retention times and in the table “Metabolites” any metabolite, e.g. NPH - Uric acid.

    NPH Uric acid NPH test mix

  2. Click Export MS instrument methods.

    Export ms instrument method

  3. In the pop-up, define the Window extension in seconds. The window extension will extend the time on either side of the WebIDQ defined integration by the specified amount for the exported MS instrument method.

    Define integration window

Recommendation: 8 sec.

  1. Click Export.
  1. Acquisition windows for the instrument method will be downloaded, e.g.
    MS instrument method KitName-0-xxxx MS vendor software (8 sec).xxx

Refer to Technical guide - RT adjustment for updating instrument methods with downloaded acquisition windows.

Integration windows features

Integration window information is shown in tables Wells and Metabolites in a specific column Info.

Integration window status "Wells" Integration window status "Metabolites"

For integration window status information details, hover with the mouse over an integration window information.

Integration window info Integration window info

  • gray, missing chromatogram(s)
  • ochre, missing integration window(s)
  • blue, default integration window in use
  • red, removed integration windows in use
  • purple, RT windows from auto peak picking with low confidence value
  • no status, specific integration window in use
    Integration window specific

Default integration window = same integration window for all samples of a plate run

To find integration window information, define a filter.

Integration window low confidence

Integration windows of low confidence, highlighted in purple, need to be checked manually.

Refer to section Check integration windows.

Chromatogram window features

To adjust the peak integration, several options are available in the Chromatograms panel. Depending if one or multiple injections/wells are selected in the table wells and the position of the zoom slider, adjustments are either applied to one or multiple chromatograms.

Applied to one chromatogramApplied to multiple chromatograms
One well selectedAll wells selected
Zoom toggle switched to 1, 2, 3, 4, or 5
e.g. Zoom 1
Zoom toggle switched to All
Zoom All
One chromatogramAll chromatograms

Quantification chromatogram with legend

1 peak integration options

2 peak display options
3 chromatogram illustration options

Peak integration options

Peak integration options

Find peakReset to default integration windowRemove integration window
Find peakReset to default integration windowRemove integration window
Perform "Auto peak picking"Reset to Default integration windowApply no peak integration

Peak integration options

Defines peak integration options.

  • Valley to valley Peak integrated from left to right curve minimum.
    Valley to valley
  • Horizontal Peak integrated from left to right curve minimum down to baseline.
    Horizontal
  • Manual User defined peak integration width.
    Manual

Peak integration options

To apply manual changes of peak integration windows, click Reintegrate or press Enter.

Reintegrate

Peak display options

Displaying of peaks can be adjusted by one of these options.

Peak display options

Peak optionExplanation
Zoom in / Zoom outincrease or decrease peak size
Autoscaleshows all peaks - for metabolites "Reset axes" is recommended
Reset axesbest view of selected peak
Zoomzoom in or out using the mouse pointer
Panmove view with mouse pointer
Download plot as a pngsave view as PNG file

Examples

Zoom in

Zoom in

Zoom in can be performed by clicking and draging a box to zoom on that region of the chromatogram or by clicking and dragging across the chromatogram horizontally or vertically to zoom on just the x- or y-axis, respectively.

Zoom out

Zoom out

Zoom out can be done by a double left-click.

Reset axes

Reset axes

Pan

Pan

Chromatogram illustration options

Switch between different illustration options by selecting samples from the table wells, selecting a metabolite from the table metabolites, and using the zoom slider.

Table "wells"Table "metabolites"Zoom
Table "wells"Table "metabolites"Zoom toggle 1

Overlay chromatograms | all wells

Overlay all chromatograms of a plate run.

Table "wells"Table "metabolites"Zoom
Table "wells"Table "metabolites"Zoom toggle All

Chromatogram view all

If all wells of a plate run are selected, integration windows are defined as "plate default integration windows".

Default mode is shown.
Default integration windows are displayed as solid lines.
Default integration window.

Default integration window = same integration window for all samples of a plate run.

Overlay chromatograms | multiple wells

Overlay multiple chromatograms of a plate run.

Table "wells"Table "metabolites"Zoom toggle
Table "wells"Table "metabolites"Zoom toggle All

Chromatogram view multiple

Show individual chromatograms | one well

Show chromatograms for the selected metabolite and its corresponding internal standards (ISTD) of the selected well.

Table "wells"Table "metabolites"Zoom toggle
Table "wells"Table "metabolites"Zoom toggle All

Chromatogram view multiple

Show individual chromatograms | two wells

Show chromatograms for the selected metabolite and its corresponding internal standards (ISTD) of the selected wells.

Table "wells"Table "metabolites"Zoom toggle
Table "wells"Table "metabolites"Zoom toggle All

Chromatogram view multiple

Three, four, and five wells can be shown together accordingly.

Show TIC

Show the TIC for a selected well.

  1. Select a well and any metabolite.

    Table "wells"Table "metabolites"
    Table "wells"Table "metabolites"
  2. Activate Show TIC.

    Show TIC

Align RT shifts

If RT shifts occurred during a plate run, to overlay all chromatograms of one metabolite activate Align RT shifts.

Example for "SDMA".

No "align RT shifts"With "align RT shifts"
Align RT shifts offAlign RT shifts on

The following options are accessible through the menu icon:

Chromatogram options |

Align ISTD RT axes

Aligns chromatograms for analyte and internal standard when retentiom times differ. Example GUDCA:

Align ISTD RT axes

Show ISTDs

Show or hide ISTD chromatogram.

Show legend

Show or hide the legend.