Skip to main content

Validation | workflow

The kit run performance is evaluated in the Validation section.

Validation

To protect validation status and results from unwanted changes, use Approve plate run to lock results.

Validate plate run

To assess the performance of a kit run, click Validate.

Validate.

Wait until the validation is completed.

The progress is indicated as a blue bar at the upper left edge of the screen.
Progress bar

Validation features

Within the validation section, the following features are available.

Summary

  • Plate run: Displays current plate run barcode. Click to load the corresponding data directly in Results.

  • Run time: Date of MS data uploaded

  • OP: Plate run OP

  • Calibration: Plate run calibration status (LC only)

    When linear extrapolation < LLOQ was applied, Calibrated with lin. est. < LLOQ is shown.

  • Plate validation: "OK" or "Not OK"

    • OK: Plate successfully passed the validation criteria.
    • Not OK: Plate did not fully meet the validation criteria. Please review "Analyte results" window.
  • Note: Click Note to open a summary of all validation notes, including a summary of excluded metabolites or wells.

    Validation summary

LOD algorithm

The limit of detection (LOD) for each metabolite within a kit run is calculated from the concentrations in the zero samples based on either the standard deviation (StdDev-based) or median (Median-based).

  • StdDev-based: LOD is calculated as median concentration of the zero samples concentrations plus three times the standard deviation of the zero samples concentrations.
    LOD = median background signal + 3 × standard deviation
  • Median-based: LOD is calculated as three times the median concentration of the zero samples.
    LOD = median background signal × 3

Requirement for StdDev-based algorithm:

At least two replicates of a zero sample of the same material per plate run, e.g. two PBS replicates.

If StdDev-based or Median-based are defined as default in the Settings > LOD algorithm, this overrules the predefined LOD algorithm of the method (OP).

Default LOD algorithm

The default LOD algorithm to use for every kit can be defined in the Settings, refer to section LOD algorithm.

Method default: LOD is calculated according to default settings in the method's OP, "StdDev-based" or "Median-based".

StdDev-based: this algorithm is always used.
Median-based: this algorithm is always used.

For validated plate runs, the LOD algorithm used is shown, e.g. "Median-based".

Used LOD algorithm
When repeating the Validation of a validated plate run and the default LOD algorithm is different from the one previously used, a warning is shown.

Example

LOD algorithm warning

  • Current LOD algorithm of plate run is "Median-based"
  • Default LOD algorithm is "StdDev-based"
  • Select whether current algorithm ("Median-based") or new ("StdDev-based") algorithm is used for plate run validation.

LOD evaluation

For LOD evaluation, two options are available and used.

  1. LOD (calc.), the LOD is calculated using zero sample(s) from the corresponding plate run and the LOD algorithm defined in WebIDQ, see LOD algorithm.
  2. LOD (from OP), LOD values based on kit validation experiments are used, available in kit document "analytical specifications". These LOD values serve as instrument specific, “minimum” LOD values.

When is “LOD (from OP)” used? If a LOD value could not be calculated for a metabolite, e.g. no noise on corresponding MRM integrated, “LOD (from OP)” instead of "LOD (calc.)" is used.

Example for metabolite “C2”

  • LOD (calc.) is “0” | blue arrow.
  • LOD (from OP) is “0.133” | green arrow.
  • For LOD evaluation, “LOD (from OP) = 0.133” is used.
  • Concentration “0.065” is flagged as “< LOD” | green arrow.

LOD evaluation example

Information: “LOD (from OP)” and "LOD (calc.)" are available in Results > Data.

LOQ definition

The limit of quantification (LOQ) is defined as

LOQ = LOD × 3.3 .

Link sample material with zero samples

If zero samples of different materials are available on one plate run, sample materials need to be linked. Please find an example below.

Zero samples

  • PBS
  • Isoprop_H2O 80_20

Sample materials

  • EDTA plasma
  • liver tissue

EDTA plasma is linked with PBS by dragging EDTA plasma to PBS.

Link plasma with PBS

liver tissue is linked with Isoprop_H2O 80_20 by dragging liver tissue to Isoprop_H2O 80_20.

Link tissue with iProp

This definitions are used for limit of detection (LOD) calculations.

  • The zero sample PBS is used for LOD calculations of EDTA plasma samples.
  • The zero sample Isoprop_H2O 80_20 is used for LOD calculations of liver tissue samples.

    Linked materials for LOD calculations tissue with iProp

Plate must be validated again after changing zero samples for new the LOD(s) to be applied

Plate view

To assess the performance of a plate run, select a QC sample from the Plate view.

The size and level of detail in the plate view is adjustable.

Adjust plate view size

Plate view S small size

Plate view Id sample IDs shown

Plate view B sample barcodes shown

Plate view L large size

  1. Select a sample, e.g. QC2 on position E2, by clicking on the corresponding well.

    Plate view

  2. Validation results are displayed in the tables Analyte results and ISTD results.

    Analyte and ISTD results table

    Coefficient of variations (CVs) are calculated as percentages when at least three QC well replicates are available per plate run.
    CV = standard deviation / mean concentration * 100

    Accuracies are calculated for quantitative metabolites only, analyte classification T1.

    • Accuracies are calculated for metabolites of analyte classification "T1"
    • CVs are calculated for metabolites of analyte classification "T1" or "T2"

  3. Validation results are illustrated in the Analyte results and ISTD results graphics.

    Analyte results graphics

    CVs are shown as CV bullet.

    Accuracies are shown as Accuracy cross.

    Acceptance ranges are shown as Acceptance ranges in the corresponding color of CV bullet or Accuracy cross.

    ISTD results graphics

    ISTD intensities are shown as ISTD cross

    Acceptance ranges are shown as Acceptance ranges

  4. Concentrations of samples and counts of internal standard (ISTD) are validated based on acceptance range criteria, resulting in defined validation statuses.
  • QC status: CV out of range and Accuracy out of range are available.
  • Calibration standard status: Accuracy out of range is available.
  • 7-point calibrated metabolites quantification range: LLOQ and ULOQ are the lowest and highest enabled calibration standard level.
    Calibration standards can be enabled and disabled, see Disable calibration standard level
    Example: Cal2 - Cal7 are enabled. LLOQ = concentration of Cal2, ULOQ = concentration of Cal7.

Chromatograms

Chromatograms to a corresponding well can be displayed.

  • Click on one well, e.g. QC2 on position E2.
  • Select an analyte, e.g. Ala.

QC2, well E2 QC2, Ala QC2, Ala chromatogram

Analyte classification

  • Results from all biocrates kits are quantitative. The default unit is µmol/L.
  • Defined metabolites are quantified using a seven-point calibration based on the calibration standards. The calibration is performed in Quantification > Calibration.
  • Other metabolites are quantified using a one-point calibration based on internal standards.
  • For details, refer to the kit specific documents "Analytical specifications".

General analyte classification

Analyte classifications are kit specific.

document Analyte specifications on Kit files

Analyte classificationInformation
Type 1 (T1)Accuracy checked during kit validation. Reproducible (precise) results with low CVs and good accuracy (accurate).
Type 2 (T2)Reproducible (precise) results with low CVs.
QualitativeAcceptance criteria for precission (CV) not met.
Not validated (NV)Validation not possible, e.g. metabolite not available in validation matrix.

Kit specific analyte classification

Analyte classifications are kit specific, described in the kit specific document Analyte specifications, section "analyte classification".

The analyte sepecification document is provided with the Kit files.

Validation statuses

  • All validation statuses like <LOD, <LLOQ, or >ULOQ refer to the measured concentrations.
  • "Measured concentrations" are values, to which no calculations were applied, e.g. data normalization or sample volume different from 10 µL.
  • When applying any calculation or normalization to results, concentrations will change but the validation status stays the same.
  • Sample material specific LOD: When using zero samples of different materials, e.g. "serum" and "tissue", material specific LOD values are used, according to section Link sample material with zero samples.
  • Summary: Validation statuses like <LOD, <LLOQ, or >ULOQ are given before calculation or normalization procedures are performed.

Example

  • Measured concentration below the LOD received the status &lt; LOD.

  • Data normalization or other calculations may change the reported concentration in Results, but not the validation status.

  • Summary: After calculation or normalization procedures were performed, concentrations may be flagged as &lt; LOD but are shown above the LOD.

    Values in bold: concentration and status do not match. After normalization, concentration of sample A is shown above the LOD but the status remains &lt; LOD.

    Sample AStatusSample BStatus
    Measured concentration0.122&lt; LOD0.198&lt; LLOQ
    Normalized concentration0.433&lt; LOD0.322&lt; LLOQ

    LOD: 0.150

    LLOQ: 0.200
    Unit: µmol/L

Analyte results

Validation status of concentration.

Available calibration standard, zero, QC, and unknown sample types.

Valid concentration is within the quantification range, see kit specific document "Analytical specifications"

&lt; LOD concentration is below the limit of detection (LOD)

&lt; threshold concentration is below the defined "height threshold" (FIA data only)

&lt; LLOQ concentration is below the lower limit of quantification (LLOQ), but above the LOD

&gt; ULOQ concentration is above the upper limit of quantification (ULOQ)

CV out of range CV is outside the acceptance range

Accuracy out of range accuracy is outside the acceptance range

ISTD out of range intensity of the internal standard (ISTD) is outside the acceptance range

ISTD results

Validation status of internal standard intensity.

Available for all sample types.

Valid intensity is within the acceptance range

ISTD out of range intensity of the internal standard (ISTD) is outside the acceptance range

Blank

ISTD signals (MRMs) are checked exclusively.

Valid below upper intensity (cps) acceptance limit

ISTD out of range if above intensity (cps) acceptance limit

Approve plate run

Approve plate run

Approve plate run requires WebIDQ core+.

  • To protect validation statuses and results from unwanted changes, click Approve plate run.
    Approve plate run
    WebIDQ functionalities which would result in integration, validation status, or results changes are now disabled.
  • To allow changes of a plate run again, click Unapprove.
    Approve plate run

Enable or disable the functionality approve plate run in the settings.

This is a global setting, affecting all users.

Settings - Approve plate runs active

Disabling the setting "Approve plate runs" resets all approved plate runs to unapproved!

Settings - Approve plate runs deactivated

Changing the option "Approve plate runs" in the settings requires the WebIDQ user role admin.

WebIDQ user role
Change WebIDQ roles in mybiocrates > users, see mybiocrates manual section Edit users.