Skip to main content

Validation | features

Quality assessment

For an easy quality assessment of a kit run, validation statuses are given on metabolite, well/injection, and plate run levels.

Classification levels

  • Metabolite: a calculated concentration for a metabolite is evaluated.
  • Injection: all concentrations of an injection are evaluated.
  • Well: all injections from one well are evaluated. If only one injection per well was performed, injection level = well level.
  • Plate run: all wells of one plate run are evaluated.

Custom QC: "QC" and "Custom QC" samples are grouped under the term "QC".

Validation criteria are only given if not described in section Validation statuses.

Metabolite level

Calibration standard acceptance criteria

  • Valid accuracy within kit acceptance criteria, in general 70-130%
  • Accuracy out of range accuracy not within kit acceptance criteria, in general 70-130%

QC sample acceptance criteria

  • Accuracy calculated exclusively for metabolites of analyte classification "T1".
  • CV calculated exclusively for metabolites of QCs and of analyte classification "T1" or "T2".
    Requirement: at least three QC replicates per plate run and concentration above LOQ.
  • Valid accuracy within kit acceptance criteria, in general 70-130%, and CV below 20% (analyte classification T1) or below 30% (analyte classification T2)
  • Accuracy out of range accuracy not within kit acceptance criteria, in general 70-130%
  • CV out of range CV above 20% (analyte classification T1) or above 30% (analyte classification T2)

Well/injection level

Well/injection status is Valid, if

at least 85% metabolites per well/injection with an analyte classification T1 are Valid and
at least 80% metabolites per well/injection with an analyte classification T2 are Valid.

Well/injection status

For this evaluation, Valid means for

  • T1 metabolites do not have Accuracy ot ouf range or CV out of range,
  • T2 metabolites do not have CV out of range.

Plate run level

Plate run validation requirement: at least three QC replicates of one level per plate run.

For plate run validation, QCs are exclusively used.

  • OK if at least 67% of all wells per plate run have a well status Valid
  • Not OK if less than 67% of all wells per plate run have a well status Valid

Plate validation status

Valid well status means QC samples do not have CV out of range or Accuracy out of range, as described in Well/injection level.

Exclude metabolite, well, sample type, or plate run

A metabolite, sample, well, or plate run can be excluded from analysis. This may be required for issues during preparation or data collection, e.g. when a sample was pipetted incorrectly onto one well position or the incorrect retention time (RT) was defined for one metabolite in the corresponding acquisition method.

Exclude a metabolite

  1. Select a well, e.g. QC2 on position E2.

    QC2, well E2

  2. Right click on a metabolite, e.g. Ala.

    Exclude Ala

  3. Add a note (e.g. "Incorrect RT defined") and click Save.

    Exclude metabolite

Multiple metabolites can be selected. Click in the field Metabolite. Make a selection, and click Save.

Exclude multiple metabolite

Exclude a well

  1. Right click on a well, e.g. QC2 on position E2.

    Exclude QC2

  2. Add a note (e.g. "Incorrect pipetting") and click Save.

    Exclude well

  • Multiple wells can be selected. Click in the field Well. Make a selection, and click Save.

    Exclude multiple wells

  • Samples from the same type (e.g. all QCs) are selected by selecting QCs.

    Exclude sample type

Exclude a plate run

  1. Above the plate, click Exclude.

    Exclude plate run

  2. Add a note (e.g. "Kit run failed") and click Save.

    Exclude plate run

Add notes

Notes can be added on metabolite, well, or plate run level.

  1. Select a metabolite or well and right click, e.g. QC2 on position E2.

    Add note

  2. Add a note and click Save.

    Add note

  • Plate notes can be added, by right click on Note.

    Plate note

Multiple metabolites or wells can be selected. Click in the field Metabolite or Well, make a selection, and click Save.

Exclude multiple wells